• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/26; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364505
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment.After removal of the stent protector, while threading the device on the guide wire, a deformation of the stent was detected.The device was not used.
 
Additional Manufacturer Narrative
Combination product: yes the returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that several struts are deformed at the distal end of the stent.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the deformed struts were initially crimped on the balloon.The stent protector was not returned for analysis.A reference stent protector could not be applied over the deformed struts without bending them further.It can therefore be excluded beyond reasonable doubt that the damage was already present on delivery of the instrument.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11581909
Report Number1028232-2021-01484
Device Sequence Number11982463
Product Code NIQ
UDI-Device Identifier07640130439146
UDI-Public7640130439146
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2022
Device Model Number364505
Device Catalogue NumberSEE MODEL NO.
Device Lot Number02202456
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer05/21/2021
Initial Report FDA Received Date03/29/2021
Supplement Report FDA Received Date05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-