| Model Number |
72203012 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Mechanical Jam (2983); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had an inner blade fracture.It was reported that the device would not oscillate and had an unusual sound.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a diagnostic hysteroscopy procedure, the device was nonfunctional in activating or it stopped rotating.The device was making an unusual noise.The procedure was extended for 5 minutes.A new device was opened and used successfully.There was no patient injury.Medtronic investigation state that there was evidence that the outer blade assembly had been bent at the distal end.Inspection of the inner blade had a deformation on the distal cutting edge.
|
| |
|
Search Alerts/Recalls
|