• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MANSFIELD TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 72203012
Medical Device Problem Codes Detachment of Device or Device Component (2907); Mechanical Jam (2983); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had an inner blade fracture.It was reported that the device would not oscillate and had an unusual sound.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a diagnostic hysteroscopy procedure, the device was nonfunctional in activating or it stopped rotating.The device was making an unusual noise.The procedure was extended for 5 minutes.A new device was opened and used successfully.There was no patient injury.Medtronic investigation state that there was evidence that the outer blade assembly had been bent at the distal end.Inspection of the inner blade had a deformation on the distal cutting edge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
MDR Report Key11582213
Report Number1282497-2021-00021
Device Sequence Number4766933
Product Code HIH
UDI-Device Identifier10884521744110
UDI-Public10884521744110
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K130215
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2022
Device Model Number72203012
Device Catalogue Number72203012
Device Lot Number4894642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2021
Initial Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
-
-