MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0204
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2021-00614, 3005168196-2021-00615, 3005168196-2021-00617.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using smart coils.During the procedure, four smart coils would not advance within their respective introducer sheaths; therefore the smart coils were removed.The procedure was completed using other smart coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the first returned smart coil revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.Forceful advancement against this resistance likely contributed to the kink and fracture observed near the pusher assembly mid-joint.Further evaluation revealed that the embolization coil was detached from the pusher assembly inside the introducer sheath.This damage was incidental to the reported complaint and likely occurred due to the separation of the pusher assembly fractured segments.Evaluation of the second returned smart coil revealed offset coil winds on the proximal end of the embolization coil.If the introducer sheath is not properly aligned within the hub of the microcatheter prior to advancement, resistance may be experienced during advancement.Forceful advancement against this resistance will result in offset coil wind on the embolization coil.During the functional test, resistance was encountered while attempting to advance the pusher assembly through the introducer sheath due to the offset coil winds on the proximal end of the embolization coil, and the pusher assembly could be advanced any further.Further evaluation revealed a kink in the pusher assembly.This damage is incidental to the reported complaint and likely occurred due to forceful advancement against resistance.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-00614; 3005168196-2021-00615; 3005168196-2021-00617.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11582664
Report Number3005168196-2021-00616
Device Sequence Number17217456
Product Code HCG
UDI-Device Identifier00814548015712
UDI-Public00814548015712
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/02/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number400SMTHXSFT0204
Device Catalogue Number400SMTHXSFT0204
Device Lot NumberF83135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 03/02/2021
Supplement Date Received by Manufacturer06/09/2021
Initial Report FDA Received Date03/29/2021
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-