| Model Number |
2326060 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Laboratory user informed manufacturer that captia measles igm kit was reporting invalid assays on two kits from lot 060 due to positive control values being above the assigned isr range.No injury was reported.
|
| |
|
Search Alerts/Recalls
|