BOSTON SCIENTIFIC CORPORATION JAGWIRE; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number M00556581
Medical Device Problem Codes Break (1069); Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A jagwire guidewire was returned.Visual analysis of the device revealed evidence of a corewire break in addition, the guidewire was bent at the distal section.Based on the condition of the returned device, engineers determined that the failure modes observed were caused due to handling or manipulation of the device during the procedure or interaction with the scope.Once the wire was bent it could cause difficulty advancing the guidewire.Consequently, when the guidewire was crossing the scope extra force could have been applied leading to the corewire break.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on event which led to the reported event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.That no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagwire was used in the biliary tract during a procedure on (b)(6) 2021.During the procedure, the tip of the guidewire could not cross the scope.The procedure was completed with a new jagwire guidewire.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation results: corewire break or fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAGWIRE
Common Device Name
ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11585537
Report Number3005099803-2021-01298
Device Sequence Number17217597
Product Code OCY
UDI-Device Identifier08714729339151
UDI-Public08714729339151
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141820
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2023
Device Model NumberM00556581
Device Catalogue Number5658
Device Lot Number0025616578
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2021
Initial Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight59
-
-