| Model Number |
M00556581 |
| Medical Device Problem Codes |
Break (1069); Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A jagwire guidewire was returned.Visual analysis of the device revealed evidence of a corewire break in addition, the guidewire was bent at the distal section.Based on the condition of the returned device, engineers determined that the failure modes observed were caused due to handling or manipulation of the device during the procedure or interaction with the scope.Once the wire was bent it could cause difficulty advancing the guidewire.Consequently, when the guidewire was crossing the scope extra force could have been applied leading to the corewire break.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on event which led to the reported event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.That no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a jagwire was used in the biliary tract during a procedure on (b)(6) 2021.During the procedure, the tip of the guidewire could not cross the scope.The procedure was completed with a new jagwire guidewire.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation results: corewire break or fracture.
|
| |
|
Search Alerts/Recalls
|