DEPUY SYNTHES PRODUCTS LLC 12.4CM HEAVY DUTY LONG ATTCH; MOTOR, DRILL, ELECTRIC - ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
LONG-HD |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device was generating heat.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.Udi ¿ (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from the (b)(6) that during service and evaluation, it was determined that the attachment device was generating heat, and the etching triangle with exclamation mark was missing (illegible etch and missing label).It was further determined that the device failed pretest for labeling assessment and temperature assessment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|