Model Number ASD48A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the physician selected a 48mm gore® cardioform asd occluder to treat an atrial septal defect balloon sized to 27mm (static sizing 24mm) in a patient with deficient retro aortic and superior rims.The device was deployed, locked, and released without issue.Two to three minutes later, while assessing the device, it embolized into the right atrium.The physician's attempt to snare the device were unsuccessful and the occluder moved into the right ventricle and then to the right pulmonary artery.The patient was sent directly to surgery for device removal and surgical defect repair.
 
Additional Manufacturer Narrative
Fluoroscopic images of attempted device retrieval were returned for evaluation.The images show a large sheath next to the occluder and a gooseneck snare.The snare was unable to capture the right or left atrial eyelets or the leaflets of the occluder.The gore® cardioform asd occluder instructions for use list device embolism as a potential clinical and device adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE CARDIOFORM ASD OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 north 4th street
flagstaff AZ 86004
MDR Report Key11585763
Report Number2017233-2021-01821
Device Sequence Number20045102
Product Code MLV
UDI-Device Identifier00733132636525
UDI-Public00733132636525
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P050006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/11/2022
Device Model NumberASD48A
Device Catalogue NumberASD48A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer03/05/2021
Initial Report FDA Received Date03/29/2021
Supplement Report FDA Received Date05/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age44 YR
Patient Weight77
-
-