• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Model Number PXSLIMLAN135STR
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-00619.
 
Event Description
The patient was undergoing a coil embolization procedure in the inferior mesenteric artery (ima) using a lantern delivery microcatheter (lantern), pod packing coils (pod pcs), and pod coils.During the procedure, the physician successfully implanted a pod coil into the target location.After advancing the next pod pc into the target location, the physician began to re-position the coil by advancing and retracting it.Next, the pod pc became stuck inside the lantern when a few centimeters of the coil was advanced into the target site and a few centimeters of the coil was inside the lantern.Therefore, the lantern containing the pod pc was removed from the patient.The procedure was completed using another lantern and two additional pod pcs.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Type of Device
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key11585880
MDR Text Key242973300
Report Number3005168196-2021-00618
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00814548016641
UDI-Public00814548016641
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K152840
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/08/2023
Device Model NumberPXSLIMLAN135STR
Device Catalogue NumberPXSLIMLAN135STR
Device Lot NumberF102591
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received03/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-