Model Number 3000
Medical Device Problem Code Application Program Problem: Parameter Calculation Error (1449)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
Physician reported that while increasing the patient's night output current, a message was received noting that output current is not being delivered.Lead impedance and battery status were within normal limits.Internal data from the patient¿s generator was reviewed.The peak output current delivered for the generator was equivalent to the daytime value programmed.Generator device history record was reviewed; no unresolved non-conformances were identified, and the generator passed all quality control specifications prior to distribution.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Common Device Name
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
MDR Report Key11586360
Report Number1644487-2021-00439
Device Sequence Number4882435
Product Code LYJ
UDI-Device Identifier05425025750481
UDI-Public05425025750481
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3000
Was Device Available for Evaluation? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2021
Initial Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age14 YR
-
-