SMITH & NEPHEW, INC. SVCE REP DYONICS POWER II CONTROL UNIT; ARTHROSCOPE Back to Search Results
Model Number 72200873S
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the motor drive unit had a "hanpiece sensor fault" error.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of sensor fault was confirmed.Product failed functional testing with a sensor fault in port a only.Cause of errors is a defective electronic component on the main digital pcb.Product passed functional testing with a known good main pcb installed.The complaint investigation has concluded that the cause of overheating is a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REP DYONICS POWER II CONTROL UNIT
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11587087
Report Number1643264-2021-01212
Device Sequence Number20025444
Product Code HRX
UDI-Device Identifier03596010615411
UDI-Public03596010615411
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062849
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200873S
Device Catalogue Number72200873S
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date04/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-