Report confirmed.Evaluation determined that a portion of the leg together with the foot was detached.It was also noted that the foot was worn.Previous investigation performed under (b)(4) determined that the likely causes are debris in the collet and improper insertion of the tool.The user manual contains the following warnings do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.The preventive maintenance/service manual for the legend system specifies service intervals for devices based on the hospital usage level.The maximum specified service interval is 24 months.Device has been in use for approximately 27 months with no record of factory service during this period.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
|
Repair request initiated for device with the report of detached leg and foot.It was reported that there was no patient impact.On follow-up, it was confirmed that there was no patient or staff impact, there was no procedure delay and the procedure was completed using a backup device.It was also reported that the patient had no further issues post-surgery.
|