| Model Number |
PS10.0-50D235 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of device history record for the nail confirmed no deviations in the manufacturing process and that the finished product met all required quality inspections prior to release.A supplemental report will be submitted with device evaluation findings.
|
| |
|
Event or Problem Description
|
|
It was reported that during the explant procedure after successful limb lengthening, the nail was found to have corrosion at the telescoping junction and abnormal bone was seen in the same area upon explant.No patient impact has been noted at this time and the surgeon will continue to monitor and evaluate the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned nail revealed that the distraction rod of the nail had discoloration.This issue is currently being investigated through the capa process.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|