| Model Number |
TSM-360 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
|
| |
|
Event or Problem Description
|
|
Patient reported left side "chronic pain, swelling and shortness of breath." healthcare professional additionally reported left side capsular contracture baker grade iv.Patient also reported "cognitive issues, dizziness, insomnia and overwhelming anxiety"; these events are not device related.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|