• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ARCHITECT FREE T3 REAGENT KIT; RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ARCHITECT FREE T3 REAGENT KIT; RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE Back to Search Results
Model Number 7K63-25
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Was this device serviced by a third party? no.All available patient information were included, no additional patient information was available.3005094123-10/22/18-001-c.Product correction letters were issued on (b)(6) 2018 to all architect total t3 (tt3) and free t3 (ft3) customers, using the i1000sr, i2000, or i2000sr instruments and who received one of the impacted lots.The letter instructed the customers to prevent this interaction by either separating the architect ft3 and/or architect tt3 assays by running these tests on different instruments from the following assays [architect tsh, architect t-uptake, architect hiv ag/ab combo, architect cortisol, architect lh, architect progrp, architect rhtlv-i/ii, architect total psa, architect afp, and architect free psa, architect 25-oh vitamin d], or by performing daily maintenance on the instrument prior to performing batch testing for all ft3 and/or tt3 samples.The mechanism of carryover involves carryover of poly-l-lysine (pll) from other assays into the conjugate dispense step in the ft3 and tt3 assays through adherence of the pipetting probe.
 
Event or Problem Description
The customer observed low or erratic results for architect free t3 while processing with the reagent lot 21081ui00.There was no reported impact to patient management.
 
Additional Manufacturer Narrative
This follow up is submitted to populate fields d8 and/or h6 with data that had previously been provided in field h10.There is no change to the content of the data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT FREE T3 REAGENT KIT
Common Device Name
RADIOIMMUNOASSAY, TOTAL TRIIODOTHYRONINE
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
MDR Report Key11590027
Report Number3005094123-2021-00056
Device Sequence Number3356058
Product Code CDP
UDI-Device Identifier00380740014285
UDI-Public00380740014285
Combination Product (Y/N)N
Initial Reporter CountryNP
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/21/2021
Device Model Number7K63-25
Device Catalogue Number07K63-25
Device Lot Number21081UI00
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/23/2021
Supplement Date Received by Manufacturer05/24/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3005094123-10/22/18-001
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I1000SR MOD, 01L86-01, (B)(4); ARC I1000SR MOD, 01L86-01, (B)(4)
-
-