Model Number 2030-6535-1
Medical Device Problem Codes Fracture (1260); Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 03/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reported event: an event regarding crack/ fracture and loosening involving a hip screw was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events (crack/fracture) for the reported lot.There have been 1 other similar events (loosening) for the reported lot.Conclusion: it was reported that the patient had a revision due to shell loosening and migration, and a broken screw.The event could not be confirmed nor the exact cause be determined because insufficient information was provided.Additional information including patient details, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
It was reported that the patient's right hip was revised due to shell loosening and migration, and a broken screw.There was no known cause or contributor for the loosening or migration.There are no allegations against any other implants.A 50d cluster hole shell and torx bone screw were revised to a 58f tritanium shell with hex screws.Rep confirmed that no further information would be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5 CANCELLOUS BONE SCREW 35MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key11590094
Report Number0002249697-2021-00561
Device Sequence Number20028870
Product Code HWC
UDI-Device Identifier07613327037012
UDI-Public07613327037012
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K894124
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model Number2030-6535-1
Device Catalogue Number2030-6535-1
Device Lot NumberMLM6LL
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2021
Initial Report FDA Received Date03/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age52 YR
-
-