Model Number 1217-31-056 |
Device Problems
Device Dislodged or Dislocated (2923); Noise, Audible (3273)
|
Patient Problems
Ambulation Difficulties (2544); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/01/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that 4 weeks post surgery, the patient came back with hip problems and noise from the hip.The cause was poly liner dissociation leading to revision surgery and exchange of the poly liner.The pinnacle cup was left well fixed, only liner exchange was performed.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Visual examination of the provided x-ray image found evidence of disassociation.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|