• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN 100 W/GRIPTION 56MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN 100 W/GRIPTION 56MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-31-056
Device Problems Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Patient Problems Ambulation Difficulties (2544); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/01/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that 4 weeks post surgery, the patient came back with hip problems and noise from the hip.The cause was poly liner dissociation leading to revision surgery and exchange of the poly liner.The pinnacle cup was left well fixed, only liner exchange was performed.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Visual examination of the provided x-ray image found evidence of disassociation.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN 100 W/GRIPTION 56MM
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11590521
MDR Text Key243015294
Report Number1818910-2021-06489
Device Sequence Number1
Product Code KWA
UDI-Device Identifier10603295010227
UDI-Public10603295010227
Combination Product (y/n)N
PMA/PMN Number
K090998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1217-31-056
Device Catalogue Number121731056
Device Lot Number9591801
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/03/2021
Initial Date FDA Received03/30/2021
Supplement Dates Manufacturer Received05/12/2021
Supplement Dates FDA Received05/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALTRX NEUT 32IDX56OD; COMPETITOR HEAD (Z/B); PINN 100 W/GRIPTION 56MM; ALTRX NEUT 32IDX56OD; COMPETITOR HEAD (Z/B); PINN 100 W/GRIPTION 56MM
Patient Outcome(s) Required Intervention;
-
-