• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A2030-120
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was performed to treat a moderately calcified, mildly tortuous tibial artery that was 75% stenosed.The 3.0x12mm armada 14 balloon dilatation catheter was advanced over a command es guidewire.During inflation, the balloon did not open to its complete profile and a leak was observed from the hub.The device was replaced with a 3.0x200mm armada 14 balloon to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual, functional and scanning electron microscopy (sem) analysis was performed on the returned device.The reported inflation issue and leak were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined that the reported inflation issue and leak were due to a tear noted to the inner member.Although a cause for the tear could not be determined, there was no leak noted during preparation for use indicating that the damage was not pre-existing.Additionally, based on the evaluation of the returned unit and sem analysis, the inner member damage may be attributed to mechanical damage and may suggest that an object was inserted into the inflation port; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11592418
Report Number2024168-2021-02532
Device Sequence Number14373538
Product Code DQY
UDI-Device Identifier08717648156106
UDI-Public08717648156106
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberA2030-120
Device Catalogue NumberA2030-120
Device Lot Number0102141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2021
Initial Date Received by Manufacturer 03/10/2021
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date06/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMMAND ES GUIDEWIRE
-
-