Date sent to the fda: 04/16/2021.Additional h6 component code: g07002 ¿ reported condition not confirmed.Additional information: d9, h6.Additional h-3 summary: visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned samples determined that an empty opened foil and ten unopened samples of product code 1775g were received for evaluation.Visual inspection of the unopened samples, the swage, and the attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate date sent to the fda: 04/16/2021.Corrected information: d3, g1.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|