• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG SELOX ST 60; PACING LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG SELOX ST 60; PACING LEAD Back to Search Results
Model Number 346367
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
After an implantation period of approx.87 months, the rv lead impedance trend showed a steady increase.Furthermore an lead failure alert was reported.The patient underwent surgery for a new rv lead and the lead was capped.
 
Additional Manufacturer Narrative
As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The analysis is thus based on the inspection of the manufacturing documents of the device as well as on the provided data.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.The analysis of the provided trend data, acquired between (b)(6)2021 and (b)(6) 2021 recorded the gradually increasing pacing impedance from initially 700ohms to greater than 2000ohms.In spite of the available information, no conclusion can be drawn regarding the root cause of the clinical observation.An analysis of the lead itself would be necessary to determine the root cause.Should additional information or the device itself become available for analysis, the investigation will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELOX ST 60
Common Device Name
PACING LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key11593306
Report Number1028232-2021-01518
Device Sequence Number17022080
Product Code NVN
UDI-Device Identifier04035479077853
UDI-Public04035479077853
Combination Product (Y/N)N
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2015
Device Model Number346367
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer08/23/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-