• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 1-2 9MM LT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 1-2 9MM LT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74035221
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an inspection in the warehouse it was noticed the failure of defective devices.It was noticed that the nail prox drop ((b)(4)) has a defective attachment.Four genesis ii tibial alignment spiked fixation rod ((b)(4)), the screwdriver release handle ((b)(4)), the journey ii bcs articular insert trial size 5-6 9mm right ((b)(4)), the journey ii bcs articular insert trial size 5-6 9mm left ((b)(4)), r3 0 deg 32 mm trial 48 mm ((b)(4)), the honeycomb ((b)(4)), the box chisel ((b)(4)), journey ii bcs articular insert trial size 1-2 11 mm right ((b)(4)), the lag screw 3.2 guide pin sleeve ((b)(4)), the medium hexdriver ((b)(4)), the subtroc lag screw driver ((b)(4)), the accord dual ended hex driver ((b)(4)) two thin osteo blade 8x3 ((b)(4)), the thin osteo blade 12x3 ((b)(4)) and the journey dcf ap fem cut blk 4 ((b)(4)) were noticed cracked.The nail prox drop ((b)(4)), the meta anterior drop ((b)(4)), the mi z handle acet reamer ((b)(4)) and two thin osteo blade 8x3 ((b)(4)5) were noticed broken.No case involved.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event or Problem Description
It was reported that during an inspection in the warehouse it was noticed the failure of defective devices.The journey ii bcs articular insert trial size 5-6 9mm left was noticed to be cracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS ART ISRT TRL SZ 1-2 9MM LT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
UK  38116
MDR Report Key11593373
Report Number1020279-2021-02505
Device Sequence Number17048195
Product Code JWH
UDI-Device Identifier00885556156735
UDI-Public00885556156735
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74035221
Device Catalogue Number74035221
Device Lot Number18CB00547
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer07/08/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date07/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-