• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71173547
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2021
Event Type  malfunction  
Event Description
It was reported that during a periloc l-p tibia plate removal, the dr removed plate and screws for a patient that healed.Original surgery was in 2015 at st joseph regional medical center.During the case the screwdriver became stuck in the screwdriver handle.The case was completed as expected and no further complications were reported.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, the product analysis and the reported event could not be confirmed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE SCREWDRIVER HANDLE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11593485
MDR Text Key243095237
Report Number1020279-2021-02544
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010519580
UDI-Public03596010519580
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71173547
Device Catalogue Number71173547
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received04/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-