• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNIV ISRT SPACER SZ1-2 18M; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNIV ISRT SPACER SZ1-2 18M; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74033616
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an inspection in the warehouse it was noticed the failure of defective devices.It was reported that three universal insert spacer size 1-2 18 mm ((b)(4)), the tibial spacer block standard ((b)(4)) the journey ii xr tibial base implant impactor ((b)(4)) and the journey uni tibial trail insert size 3-4/ 9 mm ((b)(4)) were noticed cracked.The screw depth gauge ((b)(4)), meta-tan anti rotation bar ((b)(4)), humeral ruler ((b)(4)), two screw driver release handle ((b)(4)), the journey dcf ap fem cut blk 4 ((b)(4)) the journey ii tka femoral trial left size 5 ((b)(4)) and 3.5 mm locking screw guide ((b)(4)) were noticed broken.No case involved.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation.Therefore, the product analysis and the reported event could not be confirmed at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIV ISRT SPACER SZ1-2 18M
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11593489
Report Number1020279-2021-02546
Device Sequence Number3474019
Product Code JWH
UDI-Device Identifier03596010672001
UDI-Public03596010672001
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74033616
Device Catalogue Number74033616
Device Lot Number18HAP0221
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer03/05/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-