• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUMERAL RULER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUMERAL RULER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71751126
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an inspection in the warehouse it was noticed the failure of defective devices.It was reported that three universal insert spacer size 1-2 18 mm ((b)(4)), the tibial spacer block standard ((b)(6)) the journey ii xr tibial base implant impactor ((b)(4)) and the journey uni tibial trail insert size 3-4/ 9 mm ((b)(4)) were noticed cracked.The screw depth gauge ((b)(4)), meta-tan anti rotation bar ((b)(4)), humeral ruler ((b)(4)), two screw driver release handle ((b)(4)), the journey dcf ap fem cut blk 4 ((b)(4)) the journey ii tka femoral trial left size 5 ((b)(4)) and 3.5 mm locking screw guide ((b)(4)) were noticed broken.No case involved.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMERAL RULER
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
MDR Report Key11593503
Report Number1020279-2021-02559
Device Sequence Number3474025
Product Code HSB
UDI-Device Identifier03596010514714
UDI-Public03596010514714
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71751126
Device Catalogue Number71751126
Device Lot Number19CST0003
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer10/28/2021
Initial Report FDA Received Date03/30/2021
Supplement Report FDA Received Date10/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device N
Patient Sequence Number1
-
-