|
SMITH & NEPHEW, INC. MTX CEMENTED CUP X BAR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
|
Back to Search Results |
|
| Model Number |
MT2201 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the instruments broke during surgery.The procedure was successfully completed without delay using a smith+nephew back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device revealed scratches and nicks.The device was also found to be dull from excessive wear and tear.The device shows signs of extensive use.A functional evaluation of the returned device could not confirm the stated failure mode.The device functioned as intended.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|