| Model Number |
VENUM14140 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 07/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during stent deployment the stent proximal end allegedly failed to expand.The stent released after 30 seconds.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during stent deployment in the iliac vein the stent proximal end allegedly failed to expand.Rotation of the system and waiting for 30 seconds led to final stent expansion.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.Manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not available, and images demonstrating stent or delivery system were not provided.The alleged issue could not be re produced which led to an inconclusive evaluation result.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instruction for use for this product was conducted.The instruction for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: b5, d4(expiry date: 07/2022), g3.H11: h6(method, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Search Alerts/Recalls
|