| Model Number |
978A141 |
| Medical Device Problem Codes |
Break (1069); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare professional (hcp) via a manufacturer's representative (rep) regarding a patient who was undergoing an implant for an implantable neurostimulator (ins) for gastrointestinal /pelvic floor therapy.It was reported that the lead introducer top broke off the metal when removing.On march 30th the rep reported: the cause was attributed to top portion broke off during use.Most likely cause was introducer weak weld.This was replaced with introducer from another lead kit to resolve the issue.No symptoms were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the introducer revealed a procedural non-conformance.The dilator tube was separated from the hub.There were tool marks on the dilator tube.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
(b)(6) 2021, mpxr 818659, e1, rp (rep, hcp): information was received from a healthcare professional (hcp) via a manufacturer's representative (rep) regarding a patient who was undergoing an implant for an implantable neurostimulator (ins) for gastrointestinal /pelvic floor therapy.It was reported that the lead introducer top broke off the metal when removing.On march 30th the rep reported: the cause was attributed to top portion broke off during use.Most likely cause was introducer weak weld.This was replaced with introducer from another lead kit to resolve the issue.No symptoms were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The coding was changed to the introducer: continuation of d10: product id neu_pneneedle_acc lot# unknown serial# ubd: unkn, udi: unkn product type accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|