| Model Number |
72202961S |
| Medical Device Problem Codes |
Crack (1135); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).The reported device, intended for use in treatment, was received for evaluation there was a relationship found between the returned device and the reported incident.A visual inspection found distal tip and fiber damage and cracked distal lens with scratches.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during the set-up of a knee arthroscopy, the lens of the video arthroscope hd 4mm x 30 deg was cracked.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.Results of investigation have concluded that this unit had cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|