Model Number 23AGN-751
Medical Device Problem Codes Material Separation (1562); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
Related manufacturer reference number: 2648612-2021-00034.On (b)(6) 2021, a 23mm regent mechanical heart valve was selected for implant.Both leaflets of the valve became detached during typing of the valve sutures and fell into the left ventricle (lv) cavity.The valve was removed and both leaflets were retrieved from the lv; neither leaflets were reported to be fractured.A new 21mm regent mechanical heart valve was then implanted.As the physician was starting anatomy closure, one leaflet detached and fell into the lv cavity.The 21mm valve was also explanted and the leaflet was retrieved from the lv cavity; this leaflet was reported to have been fractured.A new competitor's on-x was successfully implanted.The physician alleged an increased pump time and the patient was "critical" after the procedure.Post-operatively, the patient was reported to be in stable condition.
 
Additional Manufacturer Narrative
The reported event of dislodged leaflets could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Information from the field indicated that a smaller, 21mm valve, was subsequently attempted to be implanted.This is possible evidence of oversizing.Please note, per the instructions for use, artmt600080902, rev.B, "proper valve size selection is crucial.Do not oversize the valve.If the native annulus measurement falls between two sjm regent¿ mechanical heart valve sizes, use the smaller size sjm regent¿ mechanical heart valve.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key11596135
Report Number2648612-2021-00033
Device Sequence Number14509312
Product Code LWQ
UDI-Device Identifier05414734006101
UDI-Public05414734006101
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2025
Device Model Number23AGN-751
Device Catalogue Number23AGN-751
Device Lot Number7619232
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer06/03/2021
09/22/2021
Initial Report FDA Received Date03/31/2021
Supplement Report FDA Received Date06/10/2021
09/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
-
-