Catalog Number 109990 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during treatment with a prismaflex m150 set, the return blood line became disconnected from the filter, specified as the luer connector ¿on the top of the filter¿ is disconnected from the filter¿.It was reported air was filling the blue return line exiting the top of the filter.It was reported ¿the blood splattered onto the floor and was dripping from the control device¿.The cause of the disconnection was unknown.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided picture showed that the return screwed connector was disconnected from the filter.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|