If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Investigation summary: the device was received and evaluated at the service center.The reported complaint that the canula and suction function of the device did not work, was unable to be confirmed.No defects were found with the device upon evaluation.The device was however serviced and found to be functioning according to specifications.Since the reported condition is not confirmed, the root cause for the reported failure cannot be determined.No manufacturing record evaluation is required as no manufacturer or service related issues were found with the device.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|