| Model Number |
0350800001 |
| Medical Device Problem Code |
Pumping Problem (3016)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was overpressure.The procedure was completed successfully with no medical intervention or adverse consequences.
|
| |
|
Event or Problem Description
|
|
It was reported that there was overpressure.The procedure was completed successfully with no medical intervention or adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: it was reported by the asr thomas russell fluctuating pressures randomly ref rma 12113098 salesforce case number 07182971 the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be the motor, main board, or pressure sensor.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|