STRYKER ENDOSCOPY-SAN JOSE PKG, FLOSTEADY - ARTHROSCOPY POLE PUMP; ARTHROSCOPE Back to Search Results
Model Number 0350800001
Medical Device Problem Code Pumping Problem (3016)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was overpressure.The procedure was completed successfully with no medical intervention or adverse consequences.
 
Event or Problem Description
It was reported that there was overpressure.The procedure was completed successfully with no medical intervention or adverse consequences.
 
Additional Manufacturer Narrative
Alleged failure: it was reported by the asr thomas russell fluctuating pressures randomly ref rma 12113098 salesforce case number 07182971 the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be the motor, main board, or pressure sensor.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, FLOSTEADY - ARTHROSCOPY POLE PUMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11597571
Report Number0002936485-2021-00185
Device Sequence Number19672222
Product Code HRX
UDI-Device Identifier07613327057522
UDI-Public07613327057522
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K062380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0350800001
Device Catalogue Number0350800001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2021
Initial Date Received by Manufacturer 03/04/2021
Supplement Date Received by Manufacturer03/04/2021
Initial Report FDA Received Date03/31/2021
Supplement Report FDA Received Date05/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-