• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 9X 20MM BC IF SCRW, VENTED; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 9X 20MM BC IF SCRW, VENTED; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number Ø9X 20MM BC IF SCRW, VENTED
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a patellar tendon allograft anterior cruciate ligament reconstruction surgery the device (ar-4020c-09, lot 11764659) broke inside the patient in the middle of implanting in a 9 mm diameter tibial tunnel.The surgeon used a tap before inserting the device as a bone tendon bone graft.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint is confirmed.No device was returned for investigation, however a picture of the screw used in the procedure was received.Based on the image provided, it can be seen that the screw broke approximately after the third thread from the proximal end.As the device was not returned for investigation, the cause cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9X 20MM BC IF SCRW, VENTED
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key11599597
Report Number1220246-2021-02812
Device Sequence Number19829572
Product Code MAI
UDI-Device Identifier00888867276970
UDI-Public00888867276970
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K180662
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberØ9X 20MM BC IF SCRW, VENTED
Device Catalogue NumberAR-4020C-09
Device Lot Number11764659
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/10/2021
Supplement Date Received by Manufacturer03/10/2021
Initial Report FDA Received Date03/31/2021
Supplement Report FDA Received Date09/17/2021
Patient Sequence Number1
-
-