| Model Number |
1007710-J |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The supera stent delivery system, referenced is filed under separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2021, a percutaneous intervention was performed on the left superficial femoral artery, heavily calcified lesion.A supera stent was successfully implanted.A second supera stent (s-60-040-120-p6, 0050461) was also deployed successfully.While removing the second supera delivery system however, resistance was met between the second supera delivery system and the steelcore guide wire (1007710-j, 0080761).The guide wire and supera delivery system were removed without further issues as a single unit.There was no adverse patient effect, no intervention or treatment required, and there was no clinically significant delay.No additional information was provided regarding this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported difficulty to remove was not tested based on device condition.There were additional noted damages to the core and the coils.Also, a noted corkscrew bend on the core of the wire.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.It may be possible that anatomical conditions (heavily calcified lesion) and the previously implanted stent contributed to a reduced clearance causing the resistance between the second supera delivery system and the steelcore guide wire; however, this could not be confirmed.The additional damages to the core and the coils are likely contributed to case circumstances.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|