MEDTRONIC HEART VALVES DIVISION CG FUTURE BAND; RING, ANNULOPLASTY Back to Search Results
Model Number 638RL26
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/03/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of this 26mm annuloplasty ring, it was explanted and replaced with an unknown product.The reason for the replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Product analysis: upon receipt at medtronic's quality laboratory, visual examination revealed that the ring appeared discolored and stained, showing evidence of blood contact.Damage was noted on the sewing cuff exposing the stiffener.Radiography showed no evidence of distortion or fractures on the ring.Conclusions: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Based on the returned device, damage to the sewing cuff was noted.When the device departed medtronic's possession, no issue was noted from the device history record review.The damage may have been caused during the implant or explant procedure.Surgeons often attempt to repair valves in lieu of replacing them in an effort to preserve the native anatomy.There are times when valve repair is attempted using a repair device and subsequent post-repair evaluation demonstrates an inadequate result.This is potentially due to, but not limited to, suboptimal anatomy, surgical technique, or inappropriate sizing, rather than a malfunction of the device.At this time the root cause of the replacement cannot be determined.D9: device available for evaluation? updated h3: device evaluated? updated h6: coding updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CG FUTURE BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11599934
Report Number2025587-2021-01017
Device Sequence Number4183480
Product Code KRH
UDI-Device Identifier00643169793514
UDI-Public00643169793514
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K061127
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2025
Device Model Number638RL26
Device Catalogue Number638RL26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2021
Supplement Date Received by Manufacturer11/19/2021
Initial Report FDA Received Date03/31/2021
Supplement Report FDA Received Date12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-