• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL VMAX ENCORE 22; CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL VMAX ENCORE 22; CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION Back to Search Results
Model Number VMAX ENCORE 22D
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vyaire file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
It is reported to vyaire medical that the vmax encore 22 experienced a burning smell upon installation of sample pump.The customer confirmed there was no patient involvement associated with the reported issue.
 
Additional Manufacturer Narrative
Result of investigation: vyaire medical was able to confirmed the reported issue through failure analysis and was able duplicate the issue.The reported failure of the vmax encore 22 unit assembly p/n 777494 s/n (b)(6) caused by a missing spacer washer.Therefor it causes a short circuit on a tracing of the analyzer control board.Also, the sample pump was replaced by the user.There was additional failures of the missing transparent cover for the power supply and the missing lock washers.These are due to insufficient reassembly after the sample pump was installed by the user.The missing nuts - nut hx dblsm 4-40thd stl* p/n 808594 on solenoid 2 - valve, solenoid, 5vdc, 3-way, 0-10psig p/n 464248 of the pressure board - pwb pressure board opt vmax m tsted p/n 775383 rev e s/n (b)(6) are the result of insufficient assembly during manufacturing.Submitted a supplemental report in accordance with 21 cfr section 803.56.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VMAX ENCORE 22
Common Device Name
CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda IL 92887
MDR Report Key11599994
Report Number2021710-2021-13640
Device Sequence Number3520906
Product Code BTY
UDI-Device Identifier10846446021417
UDI-Public(01)10846446021417(11)20190315
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K942211
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVMAX ENCORE 22D
Device Catalogue Number777405-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2021
Initial Date Received by Manufacturer 03/04/2021
Supplement Date Received by Manufacturer05/27/2021
Initial Report FDA Received Date03/31/2021
Supplement Report FDA Received Date06/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-