MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3544251 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
03/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: size change, reduction of the right areola size.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old female patient underwent a primary breast augmentation with a 425cc mentor memorygel breast implant and experienced postoperative right-sided rupture.The patient underwent bilateral removal and replacement with two 765cc mentor memorygel xtra breast implant prostheses (catalog #: thpx765, left serial #: (b)(4), right serial #: (b)(4)) on (b)(6) 2021 to replace the current implants with larger size implants and reduce the size of right areola.Upon explantation, the right-sided device was found to be ruptured.
|
| |
|
Additional Manufacturer Narrative
|
|
On 9-apr-2021, the suspected medical device was returned for evaluation.On 20-apr-2021, the investigation on the suspected medical device was completed.Investigation summary: mentor conducted a visual inspection, and microscopic examination of the returned device.Visual analysis of the returned device revealed a tear at the junction between the shell and patch.A microscopic examination was performed, and the cause of the tear could not be identified.It should be noted that product failure is multifactorial.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer's reference number: (b)(4).
|
| |
|