Model Number 3544251
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 03/05/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: size change, reduction of the right areola size.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old female patient underwent a primary breast augmentation with a 425cc mentor memorygel breast implant and experienced postoperative right-sided rupture.The patient underwent bilateral removal and replacement with two 765cc mentor memorygel xtra breast implant prostheses (catalog #: thpx765, left serial #: (b)(4), right serial #: (b)(4)) on (b)(6) 2021 to replace the current implants with larger size implants and reduce the size of right areola.Upon explantation, the right-sided device was found to be ruptured.
 
Additional Manufacturer Narrative
On 9-apr-2021, the suspected medical device was returned for evaluation.On 20-apr-2021, the investigation on the suspected medical device was completed.Investigation summary: mentor conducted a visual inspection, and microscopic examination of the returned device.Visual analysis of the returned device revealed a tear at the junction between the shell and patch.A microscopic examination was performed, and the cause of the tear could not be identified.It should be noted that product failure is multifactorial.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key11600061
Report Number1645337-2021-03446
Device Sequence Number14330312
Product Code FTR
UDI-Device Identifier00081317003885
UDI-Public00081317003885
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2021
Device Model Number3544251
Device Catalogue Number3544251
Device Lot Number7287216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2021
Initial Date Received by Manufacturer 03/10/2021
Supplement Date Received by Manufacturer04/09/2021
Initial Report FDA Received Date03/31/2021
Supplement Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 YR
-
-