The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-00319, 3005168196-2021-00320, 3005168196-2021-00321.
|
|
The patient was undergoing a coil embolization procedure in the posterior communicating artery (pcom) using penumbra smart coils (smart coils), a penumbra smart coil detachment handle (handle), a non-penumbra microcatheter, and a guide catheter.It should be noted that the patient's anatomy was tortuous and did not permit the guide catheter to completely support the microcatheter during advancement to the target location.During the procedure, the physician used a non-penumbra coil to frame the target location.After advancing the first smart coil (6mm x 15cm) to the target location and repositioning it, the physician experienced moderate microcatheter tip kickback while making five attempts to detach the coil using the handle.The smart coil (6mm x 15cm) was therefore manually detached and successfully implanted.While attempting to implant the second smart coil (5mm x 10cm), the physician was unable to detach the coil using the handle after eight attempts were made; therefore, the smart coil (5mm x 10cm) was manually detached and successfully implanted.After advancing the sixth smart coil (2mm x 4cm) to the target location, seven unsuccessful attempts were made to detach the coil using the handle.It was also reported that the physician attempted to manually detach the smart coil (2mm x 4cm) without success.Upon withdrawal, the physician noticed that the smart coil (2mm x 4cm) had detached resulting in mild microcatheter kickback.The detached smart coil (2mm x 4cm) was completely contained inside the microcatheter; therefore, the microcatheter containing the detached coil was removed from the patient.The last smart coil (2mm x 3cm) was then advanced to the target location and would not detach after six attempts were made with the handle and four attempts were made to manually detach the coil.It was also reported that mild microcatheter kick back occurred.On the eleventh attempt, the smart coil (2mm x 3cm) was manually detached and successfully implanted.The procedure was ended at this point.There was no report of an adverse effect to the patient.
|