| Model Number |
45-680820-L |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The investigation of the returned coil system found the implant damaged and separated from the pusher.No indication of thermal detachment was found on the pusher's heater coil.The investigation found the pusher's monofilament with a tensile break shape at the tip, which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a portal vein embolization, an embolization coil implant detached in the microcatheter during retraction.The coil was removed from the patient.There was no reported patient injury or intervention.
|
| |
|