• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SELF CENT HIP 41X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SELF CENT HIP 41X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Model Number 1035-41-000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient came for follow up visit to surgeon after done total hemi arthroplasty surgery, patient is walking but felt click during walking.By doing x ray done by surgeon , image shows self centering bipolar head seems in unlock position.If other, describe --> total hemi replacement ( self centering bipolar ), did the patient experience a post-op device malfunction? --> unknown, did the patient experience an adverse event such as infection, non-union, allergic reaction, osteoporosis, overloading, pain, degenerative diseases, bleeding or oozing? --> unknown, did the patient require revision surgery or hardware removal? --> unknown, patient status/ outcome / consequences --> unknown, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown, (b)(4).Device property of -->none, device in possession of -->none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.--> true.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that there was no surgical delay.Affected side was the right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Visual examination of the provided photographs and x-ray images confirmed the reported event.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Visual examination of the provided photographs and x-ray images confirmed the reported event.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Visual examination of the provided photographs and x-ray images confirmed the reported event.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.,a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 41X28 RUST
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581
6107428552
MDR Report Key11601856
Report Number1818910-2021-06749
Device Sequence Number3460652
Product Code KWY
UDI-Device Identifier10603295003311
UDI-Public10603295003311
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K812672
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number1035-41-000
Device Catalogue Number103541000
Device Lot NumberJ27R13
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/11/2021
Supplement Date Received by Manufacturer04/13/2021
05/17/2021
02/28/2022
04/01/2022
Initial Report FDA Received Date04/01/2021
Supplement Report FDA Received Date04/16/2021
05/18/2021
03/02/2022
04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
C STEM VOID CENTALIZER 10 DM; C-STEM SZ1; ELITE MODULAR HD 28MM -3; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-