Model Number M0068407170
Medical Device Problem Codes Break (1069); Overheating of Device (1437); Electrical Power Problem (2925)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
Note: this report pertains to the third of three devices that were used in the same procedure.Refer to manufacturer report number mfr.Report # 3005099803-2021-01244 for the lithoclast ultrasound probe, mfr report #3005099803-2021-01245 for the lithoclast ultrasound probe and mfr report for the lithoclast ultrasound probe.It was reported to boston scientific corporation that three lithoclast ultrasound probes were used in a percutaneous nephrolithotomy procedure in the kidney performed on (b)(6) 2021.During the procedure, the three ultrasound probes broke near the connection site and the handpiece was hot.The probes worked effectively at the beginning and as they were continually used they became weaker and eventually broke.It was reported 350 millimeters of the probe detached inside the patient.The fragments were removed using reusable graspers.By the third probe, the physician noticed the break prior to the full separation of the probe from the handpiece.The procedure was completed with a fourth lithoclast ultrasound probe.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be stable.
 
Event or Problem Description
Note: this report pertains to the third of three devices that were used in the same procedure.Refer to manufacturer report number mfr.Report # 3005099803-2021-01244 for the lithoclast ultrasound probe, mfr report #3005099803-2021-01245 for the lithoclast ultrasound probe and mfr report #3005099803-2021-01247 for the lithoclast ultrasound probe.It was reported to boston scientific corporation that three lithoclast ultrasound probes were used in a percutaneous nephrolithotomy procedure in the kidney performed on (b)(6) 2021.During the procedure, the three ultrasound probes broke near the connection site and the handpiece was hot.The probes worked effectively at the beginning and as they were continually used they became weaker and eventually broke.It was reported 350 millimeters of the probe detached inside the patient.The fragments were removed using reusable graspers.By the third probe, the physician noticed the break prior to the full separation of the probe from the handpiece.The procedure was completed with a fourth lithoclast ultrasound probe.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be stable.Additional information received on april 1, 2021: during the procedure, the three ultrasound probes broke near the connection site and the handpiece was hot.The probes worked effectively at the beginning and as they were continually used they became weaker and eventually broke.The probes broke at their proximal end near the junction with the handpiece (and near the proximal end of the nephoscope).The piece that broke off of each probe was approximately 35 cm long.It was reported that there was still plenty of length outside the patient for the physician to grab and remove the entire length of the broken piece.By the third probe, the physician noticed the break prior to the full separation of the probe from the handpiece.The procedure was completed with a fourth lithoclast ultrasound probe.
 
Additional Manufacturer Narrative
The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASOUND PROBE
Common Device Name
LITHOTRIPTOR, ELECTRO-HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11602011
Report Number3005099803-2021-01247
Device Sequence Number11813773
Product Code FFK
UDI-Device Identifier08714729516613
UDI-Public08714729516613
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K012445
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2023
Device Model NumberM0068407170
Device Catalogue Number840-717
Device Lot Number0026187849
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer04/01/2021
Initial Report FDA Received Date04/01/2021
Supplement Report FDA Received Date04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-