ARTHREX, INC. 10.5 MM FLIP CUTTER II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Model Number |
10.5 MM FLIP CUTTER II |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an acl procedure, at the moment that the ar-1204af-105, makes contact with the tibial bone, the tip of the ar-1204af-105, broke within the joint space.The fragment was removed and the case was completed by using a new ar-1204af-105 without further issue.
|
| |
|
Additional Manufacturer Narrative
|
|
- it was reported that the device broke during the procedure.- visual evaluation identified that the cutter tip had fractured off the distal end of the device.However, without the return of the device, further investigation cannot be performed.Additionally, patient bone quality and surgical technique were not reported.- the root cause of the reported failure mode is undetermined.However, the most likely probable cause is applying excessive mechanical forces while prying / leveraging the device.- the reported event is confirmed based on the investigation of the returned pictures.
|
| |
|