| Model Number |
SA60WF |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the product was not returned for analysis.Product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse has reported that during five intraocular lens (iol) implant procedures, and after the insertion of the iol, debris was noted.The debris was cleared from the patients using i/a.The lenses were checked under the microscope prior to insertion into the eye, and no debris could be seen.Only after it was inserted into the eye could it be seen.The original procedure was completed in all instances and just a few minutes of extra time was used to ensure the debris had been cleared.There are five medical device reports associate with this physician.This report is associated with iol 2 of 5.
|
| |
|
Additional Manufacturer Narrative
|
|
Product evaluation: the product and reported material were not returned.Two photos were provided.The distortion in the first photo made is difficult to see the reported material.An artifact is observed near the right edge.Unable to determine if the artifact is on or under the lens.The second photo has an elongated material at the 2 o¿clock position.The nature and origin of the material cannot be determined from the photo.Iol product history records were reviewed and documentation indicated the product met release criteria.A qualified cartridge was indicated.Product history records were reviewed for the potential cartridge lot numbers provided and documentation indicated the product met release criteria.The handpiece and viscoelastic used were not provided.It is unknown if qualified products were used.The product investigation could not identify a root cause for the reported complaint.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|