• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ITREVIA 7 HF-T DF-1; CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ITREVIA 7 HF-T DF-1; CRT-D Back to Search Results
Model Number 393014
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2021
Type of Reportable Event Malfunction
Event or Problem Description
Device explanted due to eri.No adverse patient events reported.Should additional information become available, it will be added to this file.
 
Additional Manufacturer Narrative
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
 
Additional Manufacturer Narrative
The device was returned and analyzed.The device memory demonstrated a normal device function while the device was implanted and in service.The inspection of the device memory revealed that an elevated current was used for anti-bradycardia pacing.This indicates that high output was programmed that resulted in a faster discharge of the battery.The bench test of the icd revealed that all therapy functions were available, especially the current used for anti-bradycardia pacing was normal.In conclusion, the device was fully functional.There was no indication of a device malfunction.Analysis of the device revealed a normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ITREVIA 7 HF-T DF-1
Common Device Name
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key11603178
Report Number1028232-2021-01592
Device Sequence Number4138950
Product Code NIK
UDI-Device Identifier04035479129309
UDI-Public04035479129309
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Model Number393014
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2021
Supplement Date Received by Manufacturer04/06/2021
12/03/2021
Initial Report FDA Received Date04/01/2021
Supplement Report FDA Received Date04/06/2021
12/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age62 YR
Patient SexMale
-
-