Model Number SHG19-100
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reer1219 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported leak was noted at the luer lock connection point.The users also found that if a 3-way valve is installed between the cannula and the infusion line, there is no longer any leakage.No other information was provided.
 
Event or Problem Description
It was reported leak was noted at the luer lock connection point.The users also found that if a 3-way valve is installed between the cannula and the infusion line, there is no longer any leakage.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged luer adapter was confirmed.The product returned for evaluation was two 19ga x 1¿ ez huber safety infusion sets.The first sample (sample 1) exhibited use residues throughout.The safety was engaged and a cap was attached to the luer adapter.The second sample (sample 2) was received in its original sealed packaging.Attempts to infuse water through both samples using a 12ml syringe revealed both to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (>15 seconds) hydraulic pressurization.Microscopic inspection of sample 1 revealed superficial stress fractures throughout the luer adapter.Microscopic inspection of sample 2 was unremarkable.Both luer adapters were inspected using an iso taper gauge and sample 1 exhibited an out of spec luer.Sample 2 was unremarkable.While no leaks were identified, the too large luer orifice exhibited by sample 1 suggested that leaking may have occurred during use with other accessories.The stress fracturing suggested that forceful insertion of a tapered object, such as a slip-fit luer accessory likely caused/contributed to the luer damage.Sample 2 was unused, undamaged and appeared to have been returned as an unused lot sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ HUBER 19G X 25MM
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11605452
Report Number3006260740-2021-01081
Device Sequence Number3261950
Product Code FPA
UDI-Device Identifier00801741089077
UDI-Public(01)00801741089077
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K071846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSHG19-100
Device Catalogue NumberSHG19-100
Device Lot NumberREER1219
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2021
Initial Date Received by Manufacturer 03/12/2021
Supplement Date Received by Manufacturer05/17/2021
Initial Report FDA Received Date04/01/2021
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-