| Model Number |
SHG19-100 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reer1219 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported leak was noted at the luer lock connection point.The users also found that if a 3-way valve is installed between the cannula and the infusion line, there is no longer any leakage.No other information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported leak was noted at the luer lock connection point.The users also found that if a 3-way valve is installed between the cannula and the infusion line, there is no longer any leakage.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged luer adapter was confirmed.The product returned for evaluation was two 19ga x 1¿ ez huber safety infusion sets.The first sample (sample 1) exhibited use residues throughout.The safety was engaged and a cap was attached to the luer adapter.The second sample (sample 2) was received in its original sealed packaging.Attempts to infuse water through both samples using a 12ml syringe revealed both to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (>15 seconds) hydraulic pressurization.Microscopic inspection of sample 1 revealed superficial stress fractures throughout the luer adapter.Microscopic inspection of sample 2 was unremarkable.Both luer adapters were inspected using an iso taper gauge and sample 1 exhibited an out of spec luer.Sample 2 was unremarkable.While no leaks were identified, the too large luer orifice exhibited by sample 1 suggested that leaking may have occurred during use with other accessories.The stress fracturing suggested that forceful insertion of a tapered object, such as a slip-fit luer accessory likely caused/contributed to the luer damage.Sample 2 was unused, undamaged and appeared to have been returned as an unused lot sample.
|
| |
|