| Model Number |
2112-21625 |
| Medical Device Problem Codes |
Material Deformation (2976); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device remains implanted.
|
| |
|
Event or Problem Description
|
|
A company representative reported two capri height expansion drivers changed the lordosis of a capri expandable cage rather than the height (as the physician intended).After seeing the change in lordosis in the x-ray, the physician attempted to decrease the height of the cage but was unable to, reporting that the cage had "stripped." the physician completed the case with a 30-minute delay and reported that, "the distraction fit was acceptable to finish the case but not ideal for the patient." this report captures the capri expandable cage.
|
| |
|
Event or Problem Description
|
|
A company representative reported two capri height expansion drivers changed the lordosis of a capri expandable cage rather than the height (as the physician intended).After seeing the change in lordosis in the x-ray, the physician attempted to decrease the height of the cage but was unable to, reporting that the cage had "stripped." the physician completed the case with a 30-minute delay and reported that, "the distraction fit was acceptable to finish the case but not ideal for the patient." this report captures the capri expandable cage.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of complaint history associated with the subject catalog number was performed and no adverse trends were observed.Device history records were not able to be reviewed as no lot number was provided.The cage remains implanted in the patient and a x-ray was provided in lieu of device return.In the x-ray, the internal expansion tower endplate is seen to be angulated.It is likely that the height expansion driver was engaged to only one of the tower rings.In order to expand the height of the cage, the height expansion driver must be properly engaged with both the lower and upper threaded rings.Engagement of only one of the rings will cause the endplate to angle instead of increasing the overall tower height.
|
| |
|