• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH DEVICE & POWER SUP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH DEVICE & POWER SUP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801281
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  Injury  
Event Description
It was reported that when the device is turned on, the message of smith & nephew was displayed on the screen.However, the device stopped working.The treatment was started with a back-up device.There was no patient harm.It took 3 hours to resume treatment.The doctor treated with gauze until the replacement.The device will be returned to jp local service for evaluation.The results shared after its completion.
 
Manufacturer Narrative
The device, was used in treatment was not returned for evaluation with all additional information provided we have not been able to establish a relationship between the device and the reported event or determine a root cause.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found other related failures.The ifu has been reviewed and contains comprehensive instructions on the safe operation and use of the device.The associated risk files contain the reported failure/harm or event.Probable root cause maybe a component failure.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew are taking further actions relating to the failure reported and continue to monitor for adverse trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH DEVICE & POWER SUP
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11609478
MDR Text Key243612809
Report Number8043484-2021-00738
Device Sequence Number1
Product Code OMP
UDI-Device Identifier05000223494476
UDI-Public5000223494476
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number66801281
Was Device Available for Evaluation? No
Date Manufacturer Received05/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-