• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 3 8MM; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 3 8MM; TEMPLATE Back to Search Results
Model Number 00584102308
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a sz 3 8mm zuk insert surface trl was broken.Incident occurred during set up or inspection; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device is chipped and has scratches/gouges, rendering it inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.B4, g1: mdr reporting contact name and address.
 
Event or Problem Description
It was reported that two articular surface provisional sz 3 8mm were broken.Incident occurred during set up or inspection; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 3 8MM
Common Device Name
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11610601
Report Number1020279-2021-02602
Device Sequence Number3280489
Product Code HWT
UDI-Device Identifier00885556619445
UDI-Public00885556619445
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00584102308
Device Catalogue Number00584102308
Device Lot Number61605097
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2021
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date04/02/2021
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-