Model Number 71440472
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a tka, the locking mechanism of the gii tib try impl impactor broke.Instruments outside patient.Procedure finished with s&n backup device.No surgical delay or injury to patient reported due this event.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The locking mechanism of the device is fractured, rendering the device inoperable.The device was manufactured in 2011 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII TIB TRY IMPL IMPACTOR
Common Device Name
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11610707
Report Number1020279-2021-02628
Device Sequence Number19648920
Product Code FSM
UDI-Device Identifier03596010329875
UDI-Public03596010329875
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71440472
Device Catalogue Number71440472
Device Lot Number11BM04216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2021
Initial Date Received by Manufacturer 03/09/2021
Supplement Date Received by Manufacturer04/12/2021
Initial Report FDA Received Date04/02/2021
Supplement Report FDA Received Date04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-