| Model Number |
15MHPJ-505 |
| Medical Device Problem Codes |
Backflow (1064); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Aortic Valve Insufficiency/ Regurgitation (4450)
|
| Date of Event |
03/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, a 15mm masters series hemodynamic plus valve was received by abbott.The valve appeared to have been explanted and had tissue retained upon arrival.The patient was reported to have discharged in stable condition.Additional information received on 30 march 2021 reported that the valve was implanted on (b)(6) 2016 and explanted on (b)(6) 2021 due to aortic insufficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant was reported due to aortic insufficiency.The investigation found that the mechanical leaflets opened and closed completely and were freely mobile.There was patchy pannus formation on the sewing cuff, but it did not impinge upon the leaflets.No acute inflammation or significant calcifications were present.Areas of chronic inflammation were seen in the sewing cuff material.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the aortic regurgitation could not be conclusively determined.
|
| |
|