• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO PEDIATRIC CATHETER WITH SUREGRIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO PEDIATRIC CATHETER WITH SUREGRIP Back to Search Results
Model Number 52808G
Medical Device Problem Codes Inadequate Instructions for Healthcare Professional (1319); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that from the pics and description, the user seems that different gauge of the catheters were into the packaging.They found this in 10 units that are not used.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be due to "artwork design incorrect or inadequate line clearance".It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labeling could not have prevented the reported failure.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that from the pics and description, the user seems that different gauge of the catheters were into the packaging.They found this in 10 units that are not used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3 GO PEDIATRIC CATHETER WITH SUREGRIP
Common Device Name
MAGIC3 GO PEDIATRIC CATHETER WITH SUREGRIP
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11611077
Report Number1018233-2021-01847
Device Sequence Number3667760
Product Code EZD
UDI-Device Identifier00801741136863
UDI-Public(01)00801741136863
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K172247
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2021
Device Model Number52808G
Device Catalogue Number52808G
Device Lot NumberJUCT0283
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2021
Supplement Date Received by Manufacturer06/14/2021
Initial Report FDA Received Date04/02/2021
Supplement Report FDA Received Date07/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-